Vui lòng dùng định danh này để trích dẫn hoặc liên kết đến tài liệu này:
https://dspace.ctu.edu.vn/jspui/handle/123456789/22197
Nhan đề: | Simultaneous determination of arsenic and mercury in water at trace levels by differential pulse anodic stripping voltammetry using a simple gold disk electrode |
Tác giả: | Vu, Hai Dang Pham, Thi Hai Yen Pham, Hong Phong Pham, Khac Duy Vu, Thi Thu Ha |
Từ khoá: | Gold disk electrode Arsenic Mercury Simultaneous detection DPASV |
Năm xuất bản: | 2019 |
Tùng thư/Số báo cáo: | Vietnam Journal of Chemistry;No 57(03) .- Page.339-342 |
Tóm tắt: | The conventional gold disk electrode was applied for simultaneous determination of As(III) and Hg(Il) in the media of KC1 0.1 M + HC1 0.1 M. The oplimal accumulation time, the repeatability of the electrode and calibration curves were determined by DPASV technique. The electrode works stably with RSD 4.01 % and 2.98 % for As(III) and Hg(II) respectively and showed a wide linearity in the concentration range of 1-60 µg/L of both heavy metal ions. The obtained LODs were 0.096 ppb for Hg(II) and 0. 307 ppb for As(III) with 420 s of accumulation time in case of simultaneous analysis. |
Định danh: | http://dspace.ctu.edu.vn/jspui/handle/123456789/22197 |
ISSN: | 2525-2321 |
Bộ sưu tập: | Vietnam Journal of Chemistry |