Please use this identifier to cite or link to this item: https://dspace.ctu.edu.vn/jspui/handle/123456789/22197
Full metadata record
DC FieldValueLanguage
dc.contributor.authorVu, Hai Dang-
dc.contributor.authorPham, Thi Hai Yen-
dc.contributor.authorPham, Hong Phong-
dc.contributor.authorPham, Khac Duy-
dc.contributor.authorVu, Thi Thu Ha-
dc.date.accessioned2020-03-06T07:30:15Z-
dc.date.available2020-03-06T07:30:15Z-
dc.date.issued2019-
dc.identifier.issn2525-2321-
dc.identifier.urihttp://dspace.ctu.edu.vn/jspui/handle/123456789/22197-
dc.description.abstractThe conventional gold disk electrode was applied for simultaneous determination of As(III) and Hg(Il) in the media of KC1 0.1 M + HC1 0.1 M. The oplimal accumulation time, the repeatability of the electrode and calibration curves were determined by DPASV technique. The electrode works stably with RSD 4.01 % and 2.98 % for As(III) and Hg(II) respectively and showed a wide linearity in the concentration range of 1-60 µg/L of both heavy metal ions. The obtained LODs were 0.096 ppb for Hg(II) and 0. 307 ppb for As(III) with 420 s of accumulation time in case of simultaneous analysis.vi_VN
dc.language.isoenvi_VN
dc.relation.ispartofseriesVietnam Journal of Chemistry;No 57(03) .- Page.339-342-
dc.subjectGold disk electrodevi_VN
dc.subjectArsenicvi_VN
dc.subjectMercuryvi_VN
dc.subjectSimultaneous detectionvi_VN
dc.subjectDPASVvi_VN
dc.titleSimultaneous determination of arsenic and mercury in water at trace levels by differential pulse anodic stripping voltammetry using a simple gold disk electrodevi_VN
dc.typeArticlevi_VN
Appears in Collections:Vietnam Journal of Chemistry