Nhan đề: | Characteristics of hydroxyapatite coating on Ti-6A1-4V substrate fabricated Via sequent H₂O₂ -oxidizing and RF-sputtering processes |
Tác giả: | Nguyên, Thi Truc Linh Phan, Dinh Tuan |
Từ khoá: | Hydroxyapatite Ti-6A1-4-V Implant Biomaterials |
Năm xuất bản: | 2020 |
Tùng thư/Số báo cáo: | Vietnam Journal of Chemistry;Vol 58, No 05 .- P.654-660 |
Tóm tắt: | ln this study, we created a hydroxyapatite (HAp) coating layer on top of the Ti-6A1-4V substrate via sequent H₂O₂-oxidizing and RF-sputtering processes and determining the effect of H₂O₂:-oxidizing state to the adhesion between HAp coating layer and Ti-6A1-4V substrate. The results showed that the H₂O₂-oxidied Ti-6A1-4V surface is rough and porous, which increases the adhesion strength of the HAp coating layer on the alloy substrate. The shear strength value of the HAp/H₂O₂-oxidized Ti-6A1-4V substrate was 69.3 MPa, significantly higher than that of the HAp/original Ti-6A1-4V substrate (12.9 MPa). The X-ray photoelectron spectroscopy (XPS) proved the denser HAp coating layer covered the TiO₂ /Ti-6A1-4V substrate, consequently effectively prevented releasing of the unwanted toxic elements from the metallic implant. |
Định danh: | https://dspace.ctu.edu.vn/jspui/handle/123456789/42457 |
ISSN: | 2525-2321 |
Bộ sưu tập: | Vietnam Journal of Chemistry |