Please use this identifier to cite or link to this item: https://dspace.ctu.edu.vn/jspui/handle/123456789/63975
Title: High performance liquid chromatography assay for determination of moxifloxacin in human plasma
Authors: Luong, Thi My Hanh
Nguyen, Thi Minh Diep
Pham, Thi Ngoc Mai
Nguyen, Xuan Truong
Keywords: Moxifloxacin
Protein precipitation
Trichloroacetic acid
HPLC method
Human plasma
Issue Date: 2019
Series/Report no.: Vietnam Journal of Science and Technology;Vol. 57, No. 03 .- P.336-343
Abstract: A simple reversed phase HPLC method with UV detection has been successfully developed and validated for determination of moxifloxacin in human plasma. The sample pretreatment involves only single-step protein precipitation with trichloroacetic acid. Moxifloxacin was measured in plasma using a validated I-IPLC method with LUV detector at 295 nm, C 18 column (25 cm x 4.5 mm, 5 µm), a mixture of phosphate buffer pH 4.0 and acetonitrile (30: 70, v/v) as mobile phase at a flow rate of 0.8 mL/min. Retention time of moxifloxacin was found to be 7.4 min. The mean recovery for the drug was obtained 97.30 % . The calibration curve was linear over the concentration range of 0.3 to 25.0 µg/mL with coefficient correlation of 0.9991. This method was successfully applied for therapeutic drug monitoring.
URI: https://dspace.ctu.edu.vn/jspui/handle/123456789/63975
ISSN: 2525-2518
Appears in Collections:Vietnam journal of science and technology

Files in This Item:
File Description SizeFormat 
_file_
  Restricted Access
2.68 MBAdobe PDF
Your IP: 18.221.249.198


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.