Nhan đề: Research and development of a luminance responsivity standard system at the Vietnam Metrology Institute
Tác giả: Cao, Xuan Quan
Hoang, Ngoc Dung
Nguyen, Hoang Luong
Từ khoá: Luminance meter
Luminance responsivity scale (A/(cd/m²))
Luminance standard integrating sphere source
Photometry and radiometry
Năm xuất bản: 2020
Tùng thư/Số báo cáo: Vietnam Journal of Science, Technology and Engineering;Vol. 62, No. 02 .- P.03-06
Tóm tắt: The development of a luminance responsivity standard is central to the study of photometry and radiometry at the Vietnam Metrology Institute (VMI). Today, two methods are used to determine luminance responsivity. The first method uses an integrating sphere light source, also known as a luminous intensity standard lamp, and the second method compares the luminance responsivity value to that of a known standard photometer. Using these methods, it is possible to identify small uncertainties. In this work, we analysed the luminance responsivity standard system (model: VMI-PR-006) based on the national standard system of luminance (model: V11.04.18). This system was used to calibrate the luminance responsivity scale of standard photometer and luminance standard integrating sphere.