Title: Xác định hàm lượng benzalkonium clorid trong chế phẩm dung dịch xịt mũi bằng kỹ thuật HPLC
Authors: Lê, Đình Chi
Phạm, Thị Hiền
Dương, Minh Tân
Lê, Quang Thảo
Keywords: Benzalkonium chloride
C18 column
HPLC
Nasal spray
Issue Date: 2018
Series/Report no.: Tạp chí Dược học;Số 512 .- Tr.80-83
Abstract: An HPLC method was developed for assay of benzalkonium chloride in nasal sprays, as: Column - Phenomenex C18 (250 x 4.6 mm, 5 pm); Column t° - 40°C; Detector - UV(210 nm); Mobile phase - a mixture of acetonitril and 10 mM solution of potassium dihydrogenphosphate pH 2.2 containing 0.3 % (v/v) of triethyalamlne (60:40, v/v); Flow rate - 1.2 ml/mln; Injection volume - 25 gl. The method was validated in terms of linearity, specificity, precision and accuracy. These all showed the proposed method was suitable and reliable for intended application.